Maintenance Liquistation CSF48
226 Endress+Hauser
Clean the wetted parts as follows:
1. Release the thread adapter nut on the intake hose (item 6).
2. Turn the intake hose to the "open" position at the hose lock (item 5) and remove the hose
from above.
3. Release the air hose (item 4) and remove the dosing chamber (item 2) from the front along
with the outlet hose (item 1).
4. Open the bayonet lock (item 3) and open the dosing chamber.
5. Clean these parts (hoses, dosing chamber etc.) with water or soapsuds. Use a bottle brush if
necessary.
You can wash the dosing chamber and dosing chamber cover in a dishwasher at 60 ˚C.
6. Make sure the dosing pipe is set correctly and set the old value if necessary.
7. Reinstall the cleaned parts in reverse order.
Version with peristaltic pump
a0014004
Fig. 102: Version with peristaltic pump
1Pump tubing
2Pressure sensor
3 Suction line connection