Maintenance Liquistation CSF48
228 Endress+Hauser
Cleaning the distribution arm
a0014112
Fig. 104: Sample compartment
1 Distribution arm motor
2Outlet pipe
3 Distribution arm
Clean the distribution arm as follows:
1. Release the outlet pipe (item 2).
2. Push up the splash guard.
3. Remove the distribution arm from the front.
4. Remove the cover.
5. Clean these parts with water or soapsuds. Use a bottle brush if necessary.
6. Reinstall the cleaned parts in reverse order.
Make sure the distribution arm is seated correctly! The distribution arm must be locked as
otherwise the rotation movement could be blocked or the system might no longer
approach the bottles correctly.
12.4.3 Sample compartment
The sample compartment has a fully integrated plastic inner lining.
Clean the sample compartment as follows:
1. Remove the bottle trays and the distribution pan.
2. Remove the distribution arm.
3. Spray-clean the sample compartment with a water hose.
You can wash the PE and glass bottles in a dishwasher at 60 ˚C.