Commissioning Liquistation CSF48
72 Endress+Hauser
9.4.3 Programming for automatic sampling
Create a simple sampling program in the general overview under "Select sampling
program/New/Basic" or in the "Menu/Setup/Sampling programs/Setup program/New/Basic"
menu:
• Enter the "Program name".
• The settings from the "Basic setup" for bottle configuration and bottle volume are displayed.
• Sampling mode = "Time paced CTCV" is preset.
• Enter the "Sampling interval".
• Enter the "Sampling volume" per sample. (For version with vacuum pump, configure under
"Menu/Setup/General settings/Sampling".)
• Select "Bottle change mode" after number or time for average samples.
With the option "Bottle change after a time", you can enter the change time and bottle
synchronization (None, 1st bottle change time, 1st time of change + bottle number). The
description for this can be found in the "Bottle synchronization" section.
• In "Multiple bottles", enter the number of bottles the sample should be distributed over.
• "Start condition": immediately or time-delayed.
• "Stop condition": After program end or continuous operation.
• Pressing the "SAVE" soft key saves the program and ends data entry.
•Example: