QUICK SETUP “Commissioning” PROline Promass 80
4 Endress+Hauser
!
The display returns to the cell QUICK SETUP COMMISSION if you press the ESC key
combination during parameter interrogation. The stored parameters remain valid.
The “Commissioning” Quick Setup must be carried out before one of the Quick Setups
explained below is run.
Only units not yet configured in the current Setup are offered for selection in each
cycle. The unit for mass, volume and corrected volume is derived from the
corresponding flow unit.
The “YES” option remains visible until all the units have been configured.
“NO” is the only option displayed when no further units are available.
Only the outputs not yet configured in the current Setup are offered for selection in
each cycle.
The “YES” option remains visible until all the outputs have been parameterized.
“NO” is the only option displayed when no further outputs are available.
The “automatic parameterization of the display” option contains the following
basic
settings/factory settings:
YES: line 1 = mass flow; line 2 = totalizer 1
NO: The existing (selected) settings remain.