PROline Promass 80 10 Technical data
Endress+Hauser 85
Zero point stability (Promass A):
Zero point stability (Promass H, I):
Maximum measured error in % of reading (example: Promass 80 F, M / DN 25)
Calculation example (mass flow, liquid):
Given: Promass 80 F / DN 25, flow = 8000 kg/h
Max. measured error: ±0.15% ± [(zero point stability / measured value) x 100]% o.r.
Max. measured error
→ ±0.15% ± = ±0.161%
Density (liquid)
• After field density calibration or under reference conditions:
Promass F: ±0.0005 g/cc
Promass M, E, A, H: ±0.0010 g/cc
Promass I: ±0.0020 g/cc
DN Maximum full scale value
[kg/h] or [l/h]
Zero point stability
[kg/h] or [l/h]
1 20 0.0010
2 100 0.0050
4 450 0.0225
DN Maximum full scale value
[kg/h] or [l/h]
Zero point stability
Promass H
[kg/h] or [l/h]
Zero point stability
Promass I
[kg/h] or [l/h]
8 2000 0.20 0.20
15 6500 0.65 0.65
15 * 18000 – 1.8
25 18000 1.8 1.8
25 * 45000 – 4.5
40 45000 4.5 4.5
40 * 70000 – 7.0
50 70000 7.0 7.0
* DN 15, 25, 40 “FB” = Full bore versions of Promass I
F06-80xxxxxx-05-xx-xx-xx-000
0,9 kg/h
8000 kg/h
-------------------------- -
100 %⋅