Proline Prowirl 72 FOUNDATION Fieldbus 11 Operation via FOUNDATION Fieldbus
Endress + Hauser 93
System Unit -
Temperature
AUTO - OOS Use this function to select the desired unit for the operating temperature
to be entered in the "Process Param. – Operating Temp." function
(Page 95).
Options:
°C (Celsius)
K (Kelvin)
°F (Fahrenheit)
R (Rankine)
Factory setting:
Depends on country (metric system units, Page 121, or US units,
Page 122)
System Unit -
Length
AUTO - OOS Use this function to select the unit for the length of the nominal diameter.
The unit you select here is valid for:
Mating pipe diameter, (Page 96, "Process Param. – Mating Pipe
Diameter")
Options:
– Millimeter mm
– Inch inch
Factory setting:
Depends on country (metric system units, Page 121, or US units,
Page 122)
Process Param. -
Application
AUTO - OOS Use this function to specify the state of aggregation of the fluid.
Options:
Gas/Steam or Liquid
Factory setting:
"Liquid".
! Note!
If the option selected in this function is changed, the values in the
following functions in the Transducer Block "Display" must be adjusted if
used:
• Line 1 – 100% Value, Page 108
• Line 2 – 100% Value, Page 109
" Caution!
If the option selected in this function is changed, the totalizer is reset
to 0.
Transducer Block "Flow" (Endress+Hauser parameters) / base index 500
Parameter Write access for
operating mode
(MODE_BLK)
Description