Operating conditions (continued)
Process conditions
Nominal pressure Promag A
PN 16 for Tri-Clamp and PVC couplings
PN 40 for all other connections
Promag H
PN 16
Promag F
DIN PN 6 (DN 1200...2000)
PN 10 (DN 200...1000)
PN 16 (DN 65...150)
PN 40 (DN 15...50)
PN 10 (DN 1200...2000, optional)
PN 16/25 (DN 200...1000, optional)
PN 40 (DN 65...150, optional)
ANSI Class 150 (
1
/
2
...24")
Class 300 (
1
/
2
...6", optional)
AWWA Class D (28...48")
JIS 10K (DN 50...300)
20K (DN 15...40)
20K (DN 50...300, optional)
The material load curves (p-T-load diagrams) for all process connections
can be found in the Technical Information TI 027D/06/en “Promag 33”
Conductivity
Minimum conductivity:
5 µS/cm (for liquids in general)
20 µS/cm (for demineralised water)
With the remote “FS” version the conductivity required also depends on
the length of the cable see page 109
“Connection cable length”
Pressure loss
No pressure loss if sensor and piping have the same nominal diameter.
Pressure loss specifications when using adapters e.g. reducers or
expanders see page 14.
Vacuum resistance of measuring tube lining see page 115
Mechanical construction
Design / Dimensions
Dimensions see pages 99–105
Internal diameter of measuring tube see page 114
Weight
See pages 99–105
Materials Transmitter housing:
Powder-coated die-cast aluminium
Sensor housing:
Promag A 1.4435 incl. threaded stub
Promag H 1.4301
Promag F DN 15...300: Powder-coated die-cast aluminium
DN 350...2000: Coated steel
(continued on next page)
Promag 33 9 Technical Data
Endress+Hauser 111