Operating Matrix for Rackbus RS 485
H0 H1 H2 H3
V0
MEASURED
VALUE
FLOWRATE TOTALIZEDVOL.
FLOW RATE UNITS
0: dm3/s 12: gal/h
1: dm3/min 13: gal/day
2: dm3/h 14: gpm
3: m3/s 15: gph
4: m3/min 16: gpd
5: m3/h 17: mgd
6: l/s 18: bbl/min
7: l/min 19: bbl/h
8: l/h 20: bbl/d
9: hl/min 21: ft3/s
10: hl/h 22: cc/min
11: gal/min
VOLUME UNITS
0: dm3
1: m3
2: l
3: hl
4: gal
5: bbl
6: Kgal
7: ft3
V1
CURRENT OUTPUT
FULL SCALE 1
DUAL RANGE MODE
0: OFF
1: ON
FULL SCALE 2
ACTIVE RANGE
0: RANGE 1
1: RANGE 2
V2
PULSE / FREQ.
OUTPUT
OPERATION MODE
0: FREQUENCY
1: PULSE
PULSEVALUE PULSEWIDTH FULL-SCALE FREQ.
V3
RELAYS
RELAY 1 FUNCTION
0: ERROR
1: EPD
2: ERROR + EPD
3: DUAL RANGE
4: PRE AL BATCH
5: FLOW DIRECT.
6: LIMIT VALUE K1
SWITCH-ON PT.
RE 1
SWITCH-OFF PT.
RE 1
RELAY 2 FUNCTION
0:
1: EPD
2:
3: DUAL RANGE MODE
4: BATCHING
5: FLOW DIRECT.
6: LIMIT VALUE K2
V4
BATCHING
BATCH MODE
0: OFF
1: VOLUME
BATCH QUANTITY PRE BATCH QTY COMPENSATION QTY
V5
DISPLAY
TOTAL OVERFLOW
RESETTOTALIZER
0: NO
1: YES
DISPLAY LINE 1
0:
1: FLOWRATE
2: TOTALIZER
3:
4: BATCH QUANTITY
5: BATCH UPWARDS
6: BATCH DOWNWARD
7: BATCH COUNTER
DISPLAY LINE 2
0: OFF
1: FLOWRATE
2: TOTALIZER
3: TOT. 1 OVERFL.
4: BATCH QUANTITY
5: BATCH UPWARDS
6: BATCH DOWNWARD
7: BATCH COUNTER
V6
COMMUNICATION
INTERFACE
RS 485
RACKBUS ADDRESS
V7
SYSTEM
PARAMETER
POS. ZERO RETURN
0: OFF
1: ON
ACCESS CODE
SELF CHECK
0: OFF
1: ON
V8
PROCESSING
PARA.
LOW FLOW CUTOFF
NOISE SUPPRESSION
0: OFF
1: MODERATE
2: MEDIUM
3: HIGH
EMPTY PIPE DET.
0: OFF
1: ON
2: EMPTY
3: FULL
EPD RESPONSE TIME
0: 1 s
1: 2 s
2: 5 s
3: 10 s
4: 30 s
5: 1 min
V9
SENSOR DATA
K-FACTOR POS. K-FACTOR NEG. ZERO POINT NOMINAL DIAMETER
V10
SETUP
TAG NUMBER CHA. 1
5 Display and Operation Promag 33
44 Endress+Hauser