4.7 Connecting HART Communicator
The following connection versions are available to the user:
• Direct connection to the Promag transmitter via Terminals 26/27
• Connection via the analogue 4...20 mA cable of the current output
Note!
The measuring loop must have a minimum resistance of 250 Ω.
4.8 Connecting Commubox FXA 191 (Commuwin II software)
The following connection versions are available to the user:
• Direct connection to the Promag transmitter via Terminals 26/27
• Connection via the analogue 4...20 mA cable of the current output
Note!
• The measuring loop must have a minimum resistance of 250 Ω.
• Move the DIP switch on the Commubox to ‘HART’.
• Set the function “CURRENT SPAN” to ‘4–20 mA’ (see page 51) and the function
“PROTOCOL” to ‘HART’ (see page 69).
• When connecting up, also take into account the information given in the
documentation issued by the HART Communication Foundation.
This applies especially to HCF LIT 20: “HART, a technical summary”.