Function group
RELAYS
RELAY 1
FUNCTION
Selection of the relay function. The setting “FAILURE” can only be assigned to
relay 1, not to relay 2!.
Notes!
Note page 60 and 61 for the switching response of relay 1.
For safety reasons we recommend configuring relay 1 as the alarm output
and to define the failsafe mode of the outputs (see pages 51 and 57).
As standard the normally open contact of relay 1 is brought out. However,
the normally closed contact is also available by plugging a jumper on the
communications board (see Figure below).
6
FAILURE
Indicating system errors
(list of error messages see page 90)
EPD * Empty pipe detection (indicating an empty or
partly filled measuring tube)
ERROR + EPD * Indicating system errors or an empty/partly
filled measuring tube
DUAL RANGE MODE * Indicating active full scale value (1 or 2)
BATCH PRECONTACT * Indicating reached prebatch quantity
FLOW DIRECTION Indicating direction of flow (forward/reverse).
On unidirectional measurement, relay 1 also
switches in the negative flow direction.
LIMIT FLOWRATE 1 Indicating if preset limit value is outside range
or
Indicating an overflow (v 12,5 m/s).
* These parameters only appear on the display if the appropriate
function has already been activated.
1
2
Configuring of relay contacts (Relay 1 / 2)
Factory setting relay 1 NO contact
Factory setting relay 2 NC contact
Jumper (J)
Warning!
Danger from electrical shock!
Switch off power supply before
unscrewing the cover from the
transmitter housing.
ba009y68
Note!
NC contact
brought out
NO contact
brought out
1
2
1
2
J
J
ba009y81...82
Warning!
6 Description of Functions Promag 33
58 Endress+Hauser