Function group
PULSE / FREQ. OUTPUT
FULL SCALE
FLOW
Entry of the required full scale value for the flow rate.
Scaling always applies to both directions of flow (bidirectional), but with the
unidirectional mode, no signal is given for a negative flow.
Note!
This function is only available if the setting “FREQUENCY” is selected in the
function “OPERATION MODE” (see page 53).
6
5-digit number with floating decimal point (e.g. 6400.0 dm
3
/min)
7
UNIT ==> FLOW RATE UNIT
Unit selection see “FLOW RATE UNIT function
End freqency [%]
forward
reverse
Span
Full scale
value
Q
ba009y57
Note!
Promag 33 6 Description of Functions
Endress+Hauser 55