Function group
COMMUNICATION
START PULSE
WIDTH
Certain functions of the auxiliary input are only started via a pulsed voltage
(see page 70). In this function, you enter the minimum pulse width to be
reached by the input pulse in order that the appropriate function is activated.
This ensures that the function is not activated by transient voltage peaks
(interference pulses).
Note!
This function is only available if the Promag electronics are fitted with the
communications module “RS 485” and if the auxiliary input is activated as
well as appropriately configured.
6
max. 3-digit number: 20...100 ms
Factory setting:
20 ms
SYSTEM
CONFIG.
In this function, the actual configuration of the communications module
“RS 485” is shown:
Note!
This function is only available, if the transmitter electronics are fitted with a
RS 485 communication module.
This function can only be reconfigured after entering a special service code.
If you have problems with the existing configuration, then please contact
your E+H Service organisation.
Display:
RS 485 / CURRENT
1) 2)
RS 485 / FREQUENCY
2) 3)
AUX. INPUT / CURRENT
1) 4)
AUX. INPUT / FREQ.
3) 4)
1)
The function group “PULSE / FREQ. OUTPUT is blanked out.
2)
The matrix fields for auxiliary input are blanked out.
3)
The function group “CURRENT OUTPUT is blanked out.
4)
The matrix fields for RS 485 are blanked out.
Note!
Note!
Promag 33 6 Description of Functions
Endress+Hauser 71