Function group
SYSTEM PARAMETERS
PRESENT
SYSTEM
CONDITION
With this function system/process errors as well as status messages which
occur while measurement is in progress can be called up according to their
priority.
Error and status messages are displayed in the HOME position alternately
with the actual measurement variable. On activating the diagnosis function
(
7
) there is automatically a jump to this function. With error-free measurement
the message “S: SYSTEM WORKS NORMALLY” will appear on the display.
Note!
This function can also be selected directly through the function group
“SYSTEM PARAMETERS”.
A complete listing of all possible system/process errors and status messages
is found on page 90 ff.
6
Calling up other current errors or status messages
“+” message with higher display priority
“–” message with lower display priority
7
By pressing the diagnosis function again when a system error occurs
you can also call up error descriptions. In such cases a diagnosis
symbol (stethoscope 9 ) is shown on the display.
PREVIOUS
SYSTEM
CONDITIONS
In this function, all system/process errors and status messages that have
occurred so far are listed in chronological order (error history with max.
10 entries).
Note!
A complete list of all possible system/process errors and status messages is
given on page 90 ff.
If no error or status messages have occurred since the measuring system
was last started up then the display shows the message “S: NO ENTRY
EXISTING”.
With more than 10 entries the oldest is overwritten.
Storage of this list is volatile and is lost if there is a power supply failure.
6
Calling up other system/process errors and status messages
“+” Listing is done chronologically with the oldest, second oldest ...etc.
message.
“–” Listing is done chronologically with the latest, second latest ... etc.
message.
Note!