Function group
SENSOR DATA
MAX.
SAMPLING RATE
Display of the maximum sampling rate (= SAPS). It depends on the particular
sensor being used and is set in the factory.
Caution!
Under normal circumstances, the max. sampling rate should not be altered.
The service code number required for this is known to your E+H Service
organisation. Please contact E+H Service for any problems or questions that
have occurred.
6
max. 3-digit number with fixed decimal point: 1.0...60.0 /s (per second)
Factory setting:
dependent
on the sensor
SAMPLING
RATE
Display of the sampling rate (= SAPS). The following standard values apply for
the flowmeters:
• Promag A, F → 16.7 / s
• Promag H → 25.0 / s
Note!
• The sampling rate is usually set to the MAX. SAMPLING RATE. It should only
be altered in special cases. The service code number required for this is
known to your E+H Service organisation. Please contact E+H Service for any
problems or questions that have occurred.
• The Promag 33 measuring system is synchronised with the main power supply.
Therefore, the sampling rate entered is set to the nearest possible value or
rounded off to it.
6
max. 3-digit number with fixed decimal point:
– upper limit = depending on nominal diameter (max. 60.0 / s)
– lower limit = 1.0 / s
SERIAL NUMBER
Display ot the serial number of the sensor. The serial number is normally
entered in the factory.
6
max. 6-digit number: 1...999999
EPD
ELECTRODE
This function indicates whether the sensor is equipped with an electrode for
Empty Pipe Detection (EPD). This setting is made in the factory to suit the
sensor installed.
Note!
Empty Pipe Detection can only be activated when an EPD electrode is fitted.
6
YES – NO
Factory setting:
– With an EPD electrode as standard → YES
– Sensors without an EPD electrode → NO
Caution!
Note!
Note!
6 Description of Functions Promag 33
82 Endress+Hauser