What to do if current loop is stuck at 4mA or 20mA in Endress+Hauser Measuring Instruments?
- TtararyanSep 12, 2025
If the current loop is stuck at 4 mA or 20 mA on your Endress+Hauser Measuring Instruments, check the display for an error message. Additionally, on the current loop board, check the voltage between the end of resistor R1 closest to the jumper and ground.