2.5.4 Extreme strain on the centrifuging tubes
2.6 Safety instructions on the device
CAUTION! Risk of injury from overloaded tubes.
Note the loading limits specified by the tube manufacturer.
Only use tubes which are approved by the manufacturer for the required g-force (rcf).
NOTICE! Risk from damaged tubes.
Damaged tubes must not be used, as this could cause further damage to the device and the
accessories and loss of the samples.
Before use, visually check all of the tubes for damage.
NOTICE! Risk from open tube lids.
Open tube lids can break off during centrifugation and damage both the rotor and the
centrifuge.
Carefully seal all tube lids before centrifuging.
NOTICE! Hazard to plastic tubes from organic solvents.
The density of plastic tubes is reduced when organic solvents (e.g., phenol, chloroform) are
used, i.e. the tubes could become damaged.
Note the manufacturer’s information on the chemical resistance of the tubes.