Troubleshooting
Centrifuge 5702/5702 R/5702 RH
English (EN)
52
9.2 Error messages
If an error message appears, proceed as follows:
Remedy the fault as described in the “Solution" column.
Press the open key to clear the error message form the display.
If necessary, repeat centrifugation.
Code Problem Cause Solution
LID The lid has not been
released.
Close the lid.
Press the rotary knob start/
stop.
Open the lid using the
emergency release, if
required.
LID Lid has not been locked.
Close the lid.
Er 2 Device does not
start.
Rotor is loaded
asymmetrically.
Load the rotor
symmetrically.
Er 3 The centrifuge
decelerates without
braking.
The tubes touch the
centrifuge lid.
Check the tubes.
Switch off the device.
Switch on the device and
wait for approx. 5 min.
Repeat the run.
Er 3-0 After switching on,
the display shows
Er 3.
The tubes touch the
centrifuge lid.
Er 3-2 The centrifuge
decelerates with
braking.
Error in the electronics.
Switch off the device.
Switch on the device and
wait for approx. 5 min.
Repeat the run.
Er 3-3 The centrifuge
decelerates without
braking.
Error in the electronics.
Er 5 The centrifuge
decelerates without
braking.
Error in the lid latch.
Close the lid.
Repeat the run.
Er 5-1 – Er 5-3 Run was started.
The rotor rotates.
Error in the lid latch.
Er 6 – Er 6-6 The centrifuge
decelerates without
braking.
Error in the electronics.
Allow device to cool down.
Repeat the run.
Er 7 – Er 7-2 The centrifuge
decelerates without
braking.
The maximum speed of the
rotor is exceeded.
The actual rotational speed
of the rotor deviates from
the set value. The tolerance
is exceeded.
Check if the mains/power
supply voltage corresponds
with the technical data.