61
Technical data
Centrifuge 5702/5702 R/5702 RH
English (EN)
11.5 Application parameters
100 × g – 3000 × g
• can be set in increments of 100 × g
5702 5702 R 5702 RH
Cycle time 0 s – 99 min, unlimited
()
• Adjustable to 10 min
in increments of
0.5 s,
• from 10 min in
increments of 1 min
0 s – 99 min, unlimited
()
•Adjustable to 10 min
in increments of
0.5 s,
• from 10 min in
increments of 1 min
0 s – 99 min, unlimited
()
• Adjustable to 10 min
in increments of
0.5 s,
• from 10 min in
increments of 1 min
Temperature – -9 °C – 40°C -9 °C – 42 °C
Relative centrifugal force 100 × g – 3000 × g
•can be set in
increments of 100 × g
100 × g – 3000 × g
•can be set in
increments of 100 × g
100 × g – 3000 × g
•can be set in
increments of 100 × g
Rotational speed 100 rpm – 4 400 rpm
•can be set in
increments of
100 rpm
100 rpm – 4 400 rpm
•can be set in
increments of
100 rpm
100 rpm – 4 400 rpm
•can be set in
increments of
100 rpm
Maximum load 4 x 100 mL 4 x 100 mL 4 x 100 mL
Maximum kinetic energy 2 280 Nm 2 280 Nm 2 280 Nm
Permitted density of the
material for centrifuging
(at maximum g-force
[rcf] and/or speed [rpm]
and maximum load)
1.2 g/mL 1.2 g/mL 1.2 g/mL
Inspection obligation in
Germany
no no no
Refrigerant - 128 g (230 V)
122 g (120 V)
125 g (100 V)
132 g (230 V)
122 g (120 V)