Operation
Centrifuge 5702/5702 R/5702 RH
English (EN)
26
5.2 Switching on the centrifuge
Prerequisites
The device has been installed in accordance with the operating manual.
1. Switch the centrifuge on using the mains/power switch.
2. Press the Standby key, if required.
The display shows the parameters of the last run.
3. Press the open key to open the closed centrifuge lid.
5.3 Replacing the rotor
5.3.1 Inserting the rotor
1. Align the peg of the motor shaft.
2. Place the rotor vertically onto the motor shaft from the top.
The arrows on the rotor show the position of the groove. The pegs of the motor shaft must fit into the
groove of the rotor.
If required, lift the rotor and replace it onto the motor shaft.
3. Insert the rotor key supplied into the rotor nut.
4. Turn the rotor key clockwise until the rotor nut is firmly tightened.
5.3.2 Removing the rotor
1. Turn the rotor nut counterclockwise using the rotor key supplied.
2. Remove rotor by lifting it vertically.
NOTICE! If handled incorrectly, the rotor may fall.
The swing-bucket rotor may fall if the buckets are used as handles.
Remove the buckets before inserting and/or removing a swing-bucket rotor.
Always use both hands to carry the rotor cross.
NOTICE! Material damage due to improper rotor insertion.
The motor shaft or bearing may get damaged if the rotor falls into the motor shaft guides in an
uncontrolled manner during insertion.
Hold the rotor with both hands.
Guide the rotor onto the motor shaft.