65
Rotors, tubes and adapters
Centrifuge 5702/5702 R/5702 RH
English (EN)
12 Rotors, tubes and adapters
Please note the manufacturer’s information on the centrifugation resistance of the sample tubes used
(maximum g-force).
For ordering information, refer to the English and German version of the operating manual.
Technical data of the rotors and adapters and the order numbers of the adapters can be found in chapter
"Rotors for the Centrifuge 5702/5702 R/5702 RH" of the English version of the operating manual.
12.1 Rotor A-4-38
12.1.1 Rotor A-4-38 with 4 round buckets
Eppendorf centrifuges may only be operated with rotors that are intended for use with the
corresponding centrifuge.
Only use rotors which are marked with the name of the centrifuge (e.g., 5702).
Max. g-force: 3 000 × g
Max. speed: 4 400 rpm
Rotor A-4-38 Round bucket
5702 722.006
5702 761.001
Aerosol-tight cap
5702 721.000
Max. load per
bucket (adapter,
tube and contents):
190 g
Tube Tube Adapter Bottom shape Max. g-force
Capacity Tube diameter Max. speed
Number per
adapter/rotor
Order no.
(international)
Max. tube length
with/without
aerosol-tight cap
Radius
Micro test tube round 2 900 × g
1.5 mL – 2 mL Ø 11 mm 4 400 rpm
4/16 5702 745.006 43 mm/43 mm 13.4 cm
Micro test tube flat 2 850 × g
1.1 mL – 1.4 mL Ø 8.5 mm 4 400 rpm
5/20 5702 736.007 100 mm/100 mm 13.2 cm