4.1 Selecting the location
Select the location for the device according to the following criteria:
• Mains/power connection in accordance with the name plate.
• Minimum distance to other devices and walls: 30 cm (11.81 in).
• A resonance-free bench with a horizontal and even work surface which is designed to support the
weight of the device.
• The location has good ventilation.
• The location is protected against direct sunlight.
NOTICE! If an error occurs, the objects in the immediate proximity of the device will be
damaged.
In accordance with the recommendations of EN 61010-2-020, leave a safety clearance of
30 cm around the device during operation.
Please remove all materials and objects from this area.
NOTICE! Damage from overheating.
Do not place the device near heat sources (e.g., heating, drying cabinet).
Do not expose the device to direct sunlight.
Ensure unobstructed air circulation. Maintain a clearance of at least 30 cm (11.81 in)
around all ventilation slits.
NOTICE! Radio interference.
This device is a category A product in accordance with EN 55011. There may be disturbance
to radio reception in residential areas.
Ensure that appropriate preventive measures are taken.
The mains/power switch and cutting unit of the mains/power line must be easily accessible
during operation (e.g, residual current circuit breaker).