5.9.6 Centrifuge 5702 RH: Temperature profiles
The Centrifuge 5702 RH has a regulated heating and cooling system. This enables the exact tempering of
delicate samples.
For each rotor a special temperature profile is set up in the software. The temperature profile determines
the speed at which the rotor performs the temperature control run FastTemp. The objective is to temper the
rotor chamber and the rotor to the set temperature as quickly as possible. The temperature must be
maintained with low tolerances.
Prerequisites
• The centrifuge lid is open.
1. Press the fast temp key.
The last temperature profile that was selected is displayed.
2. Use the arrow keys to select the temperature profile for the rotor that is used.
The selected temperature profile is adopted after 5 s. The display shows the standard values again.
5.10 Switching off the centrifuge
1. Open the centrifuge lid.
The residual moisture can evaporate. The pressure on the springs is released.
2. Remove the aerosol-tight caps from the buckets.
Aerosol-tight accessories may not be stored when they are connected.
3. Switch off the centrifuge using the mains/power switch.
Display Rotor
ro F 35 Rotor F-35-30-17
ro F 24 Rotor F-45-24-11
ro F 18 Rotor F-45-18-17-Cryo
ro A4 rE Rotor A-4-38 with rectangular buckets
ro A4 ro Rotor A-4-38 with round buckets
ro A8 Rotor A-8-17
ro AL L Temperature profile for all rotors