43
Programs
Centrifuge 5702/5702 R/5702 RH
English (EN)
7Programs
You can store 2 programs each on the Centrifuge 5702 R and on the Centrifuge 5702 RH.
You can set the following parameters for each program:
Centrifugation time
Temperature
Centrifugation speed
Start of time measurement (function At set rpm)
Settings for the soft ramp
7.1 Creating and storing a program
You can store two programs on the device
Prerequisites
Rotor stop.
1. Use the rotary knob time to set the centrifugation time.
2. Use the temp arrow keys to set the temperature.
3. Use the rotary knob speed to set the speed of centrifugation.
4. Set the start of time counting (function At set rpm). Press rotary knob time > 2 s to do this.
5. Press the short key > 5 s to set the soft ramp.
6. Select the program slot. Press the prog 1 key or the prog 2 key > 2 s.
A signal sounds.
The program key does not flash any more. The program key lights up blue.
The parameters of the program are saved.
7.2 Saving the current settings as a program
You can save the current settings as a program.
Prerequisites
Rotor stop.
Press the prog 1 key or the prog 2 key > 2 s.
A signal sounds.
The program key lights up blue.
The parameters of the program are saved.