21
Installation
Centrifuge 5702/5702 R/5702 RH
English (EN)
4.3 Installing the instrument
Prerequisites
The device has been prepared in accordance with the operating manual.
The device has adapted to the ambient temperature (waiting time 3 h).
The compressor is ready for operation (waiting time 4 h).
1. Connect the mains/power cord of the centrifuge to the mains/power supply.
2. Switch on the centrifuge using the mains/power switch.
•The standby key lights up green.
•The display is active.
3. Open the centrifuge lid using the open key.
WARNING! Risk from incorrect supply voltage
Only connect the device to voltage sources which correspond to the electrical
requirements on the name plate.
Only use sockets with a protective earth (PE) conductor and suitable mains/power cord.
NOTICE! Damage to electronic components due to condensation.
Condensate can form in the device after it has been moved from a cool environment to a
warmer environment.
After installing the device, wait for at least 3 h. Only then connect the device to the mains/
power line.
NOTICE! Centrifuge 5702 R, Centrifuge 5702 RH: Compressor damage after improper
transport.
After installation, wait 4 hours before switching on the centrifuge.