Troubleshooting
Centrifuge 5702/5702 R/5702 RH
English (EN)
9 Troubleshooting
If you cannot remedy an error with the recommended measures, please contact your local Eppendorf
partner. The contact addresses can be found on the Internet at www.eppendorf.com
.
9.1 General errors
Problem Cause Solution
No display. No mains/power connection.
Check the mains/power connection.
Mains/power outage.
Check the mains/power fuse of the
device (see Replacing the fuses on
p. 49).
Check the mains/power fuse of the
laboratory.
Lid of the device
cannot be opened.
Rotor is still running.
Wait for rotor to stop.
Mains/power outage. 1. Check the mains/power fuse of the
device (see Replacing the fuses on
p. 49).
2. Check the mains/power fuse of the
laboratory.
3. Activate the emergency lid release
(see p. 55).
Device cannot be
started.
Lid of the device is not closed.
Close the lid of the device.
Device shakes
when it starts up.
Rotor is asymmetrically loaded. 1. Stop the device and load
symmetrically.
2. Restart the device.
Temperature
display flashes.
(only Centrifuge
5702 R, Centrifuge
5702 RH)
Temperature deviation from set value:
±3 °C.
Check the settings.
Check unhindered air circulation
through the air slots.
Thaw ice or switch off device and
allow it to cool down.
Standby key lights
up red.
Centrifuge not ready for operation.
Press the Standby key.