11.6 Acceleration and deceleration times
The following table contains approximate acceleration times and deceleration times in accordance with
DIN 58 970. The values are for guidance only. Fluctuations may occur depending on the condition of the
device and the load.
Rotor Times Centrifuge 5702 Centrifuge 5702 R Centrifuge 5702 RH
Soft ramp Soft ramp Soft ramp
A-4-38 with
round
bucket
Acceleration
time
19 s 1:38 min 16 s 1:37 min 16 s 1:37 min
Deceleration
time
18 s 1:31 min 22 s 1:33 min 23 s 1:32 min
Tolerance ±5 %, 5 s minimum
A-4-38 with
rectangular
buckets
Acceleration
time
18 s 1:38 min 16 s 1:37 min 16 s 1:37 min
Deceleration
time
19 s 1:30 min 22 s 1:33 min 22 s 1:32 min
Tolerance ±5 %, 5 s minimum
A-8-17 Acceleration
time
14 s 1:37 min 15 s 1:37 min 15 s 1:37 min
Deceleration
time
17 s 1:34 min 19 s 1:35 min 19 s 1:33 min
Tolerance
FA-45-24-11 Acceleration
time
13 s 1:36 min 14 s 1:36 min 15 s 1:37 min
Deceleration
time
16 s 1:32 min 19 s 1:36 min 19 s 1:32 min
Tolerance ±5 %, 5 s minimum
F-35-30-17 Acceleration
time
17 s 1:41 min 20 s 1:38 min 19 s 1:38 min
Deceleration
time
17 s 1:30 min 28 s 1:30 min 29 s 1:29 min
Tolerance ±5 %, 5 s minimum