31
Operation
Centrifuge 5702/5702 R/5702 RH
English (EN)
5.6 Closing the centrifuge lid
1. Check that the rotor is attached correctly.
2. Press the centrifuge lid down until it is gripped by the lid latch. The lid will be closed automatically.
The symbol appears on the display.
5.7 Aerosol-tight centrifugation
Aerosol-tight caps are available for the round buckets of rotor A-4-38.
5.7.1 Aerosol-tight centrifugation in a swing-bucket rotor
For aerosol-tight centrifugation in a swing-bucket rotor, use buckets with aerosol-tight caps (see Closing
round buckets with a cap on p. 30).
WARNING! Risk of injury when opening or closing the centrifuge lid.
There is a risk of crushing your fingers when opening or closing the centrifuge lid.
When opening or closing the centrifuge lid, do not reach between the lid and device or into
the latching mechanism of the lid.
Always open the centrifuge lid completely to prevent it from falling.
WARNING! Damage to health due to leaking cap.
The aerosol tightness of caps is reduced by autoclaving, mechanical strain and contamination
by chemicals.
Check the cap and the gasket before each use. Only use caps work with undamaged and
clean gaskets.
Replace the caps after 50 autoclaving cycles.
Store the cap separately. Do not screw the cap onto the bucket.
The aerosol tightness of the cap has been tested and certified in accordance with Annex AA of
IEC 1010-2-020.