Device settings
Centrifuge 5702/5702 R/5702 RH
English (EN)
6 Device settings
6.1 Changing the operating state
The centrifuge has 2 operating states, the ready state and standby mode. You can actively change between
the two operating states.
The Standby key shows the operating state of the device.
The device is ready for operation: The Standby key lights up green.
The device is in standby mode: The Standby key lights up red.
Prerequisites
The centrifuge is not performing a run.
Press the Standby key to change the operating state.
The operating state changes. The Standby key changes color.
6.2 Key lock
The display shows if the key lock is activated.
When the key lock is activated, the following centrifugation parameters cannot be changed:
Centrifugation time
Centrifuge 5702 R, Centrifuge 5702 RH: temperature
g-force and/or rotational speed
•Soft ramps
Status of the At set rpm function
The following settings can still be changed when the key lock is activated:
Starting and stopping centrifugation. Press the rotary knob time for this.
Setting the unit for the speed of centrifugation [rpm/ rcf]. Press the rotary knob speed for this.
Activating and deactivating the speaker. Press the open key > 2s.
Activating the key lock
Prerequisites
The centrifuge lid is open.
Press the short key and the open key simultaneously for > 5 s.
The centrifugation parameters cannot be changed.
Centrifuge 5702 Centrifuge 5702 R Centrifuge 5702 RH
Key lock is activated Lo on
Key lock is deactivated Lo OF