37
Operation
Centrifuge 5702/5702 R/5702 RH
English (EN)
The display shows the following information:
FA
Actual temperature in the rotor chamber
•Rotational speed
The temperature control run FastTemp stops automatically when the set temperature has been reached.
When the speaker is activated, an alarm will sound periodically.
4. Press the rotary knob time to end the temperature control run FastTemp early.
5.9.4.2 Temperature control run with aerosol-tight caps
If you perform a temperature control run and close the buckets with aerosol-tight caps, a negative pressure
will be generated in the buckets. After the temperature control run the caps cannot be removed.
1. Do not use aerosol-tight caps for temperature control runs.
2. Temper bucket and adapters without aerosol-tight caps.
5.9.5 Continuous cooling
Continuous cooling maintains the rotor chamber at the set temperature if the rotor stops.
During continuous cooling the display shows the set temperature.
To prevent the rotor chamber from freezing or condensation from forming, the temperature does not go
below 4 °C irrespective of the set temperature.
If the rotor stops, temperature control is slower than during centrifugation or a temperature control run.
Continuous cooling ends after 8 h.
Prerequisites
The centrifuge is switched on.
The centrifuge lid is closed.
The set temperature is lower than the ambient temperature.
1. Continuous cooling starts automatically.
The centrifuge only stops the temperature control run FastTemp once the rotor has
reached the set temperature. Therefore, there may be a delay between the display of the
set temperature that has been reached and the automatic end of the temperature control
run.
The set temperature can be changed during the temperature control run, using the temp
arrow keys. The duration and the speed of the temperature control run are adjusted
automatically.