25
Operation
Centrifuge 5702/5702 R/5702 RH
English (EN)
Abb. 5-3: Display Centrif uge 5702
Fig. 5-3: Display Centrifuge 5702
Abb. 5-4: Display Centrif uge 5702 R, Cen trifuge 5702 R H
Fig. 5-4: Display Centrifuge 5702 R, Centrifuge 5702 RH
1 Centrifugation time [min, s]
Actual value
2Function At set rpm
: time counting starts when 95% of the
specified g-force [rcf] or speed [rpm] has been
reached.
: time counting begins immediately.
3 Rotational speed [rpm] or g-force [rcf]
Actual value
4 Status of the centrifuge
: The centrifuge lid has been released.
: The centrifuge lid is locked.
(Flashes): Centrifugation in progress.
1 Centrifugation time [min] or [s]
Actual value
2Function At set rpm
: time counting starts when 95% of the
specified g-force [rcf] or speed [rpm] has been
reached.
: time counting begins immediately.
3Soft ramp
: Rotor accelerates and brakes slowly.
No symbol: Rotor accelerates and brakes rapidly.
4 Temperature in the rotor chamber [°C]
Actual value
5Key lock
: The key lock is activated. Parameters cannot
be changed.
: The key lock is not activated.
6Speaker
: The speaker is switched on.
7 g-force [rcf] or rotational speed [rpm]
Actual value
8 Status of the centrifuge
: The centrifuge lid has been released.
: The centrifuge lid is locked.
(Flashes): Centrifugation in progress.
Centrifuge 5702 R only: When setting the soft ramp the symbol only appears on the
display from serial number 03556.
For devices with serial numbers < 03556 (see Setting a soft ramp on p. 34).
min
sec
C
rpm
rcf
1 2
3
4
5
6
7 8