53
Troubleshooting
Centrifuge 5702/5702 R/5702 RH
English (EN)
Er 8 The centrifuge
brakes.
Error occurs when the
device accelerates or
brakes.
Repeat the run.
Er 9 – Er 9-4 The data of a run is
not saved.
Error in the electronics.
Er 10 – Er 10-5 The data of the last
run is not saved.
Error in the electronics.
Er 11 The centrifuge
decelerates without
braking.
Mains/power outage
during a run.
Error in the electronics.
Check the mains/power
cord.
Repeat the run.
Er 14 The centrifuge can
no longer be
operated.
Error in the electronics.
Restart the device.
Er 15/Inb The centrifuge
switches off and
decelerates without
braking.
Rotor is loaded
asymmetrically.
Load the rotor
symmetrically.
Repeat the run.
Er 16 – Er 16-2 The centrifuge
switches off and
decelerates without
braking.
Error in the electronics.
Check the tubes.
Switch off the device.
Switch on the device and
wait for 5 min.
Repeat the run.
Er 17 – Er 17-2 The centrifuge
switches off and
decelerates without
braking.
Error in the electronics.
Allow device to cool down.
Repeat the run.
Er 18 – Er 18-3 The centrifuge
switches off and
decelerates without
braking.
(only Centrifuge
5702 R, Centrifuge
5702 RH)
The temperature in the
rotor chamber deviates
from the set temperature
by more than 5 °C.
Check the ambient
temperature.
Check if the device is
exposed to direct sunlight.
Check if there is sufficient
clearance around the
device.
Er 19 – Er 19-1 The cooling
aggregate is
switched off.
The fan continues
to run.
(only Centrifuge
5702 R, Centrifuge
5702 RH)
Error in the cooling circuit.
Check if the air can
circulate through the
ventilation slots.
Check if there is sufficient
clearance around the
device.