Troubleshooting
Centrifuge 5702/5702 R/5702 RH
English (EN)
54
Er 20 The centrifuge
switches off and
decelerates with
braking.
The cooling
aggregate is
switched off.
(only Centrifuge
5702 R, Centrifuge
5702 RH)
Error in the electronics
Repeat the run.
Er 21 The centrifuge
switches off and
decelerates with
braking.
Error in the electronics
Er 22 The fan is switched.
The error is only
displayed if the
centrifuge does not
perform a run.
Error in the electronics.
The device can be used.
Er 23 The centrifuge
decelerates without
braking.
The ambient temperature
is too high.
The motor is too hot.
Check the ambient
temperature.
Let the motor cool down.
Repeat the run.
Er 24 – Er 24-3 The centrifuge
switches off and
decelerates with
braking.
(only Centrifuge
5702 R, Centrifuge
5702 RH)
Error in the cooling
aggregate.
Allow device to cool down.
Er 25/Int Mains/power outage
during a run.
The mains/power supply
voltage fluctuates.
The mains/power supply
voltage does not
correspond with the
technical data.
Check the mains/power
cord.
Wait until the rotor stops
rotating.
Repeat the run.
Er 27 Error in the electronics.
Repeat the run.
Code Problem Cause Solution