1 TEST PROGRAMS AND ADJUSTMENTS112 / 266
Adjustment of
remote control for Neurotest
Neurotest board
EE 30121-442
Setting the remote control for Neurotest board EE 30121-442
Jumper JP4 on the front panel is used to activate the display of the feedback frequency of output current
measurement and display of the feedback frequency of the remote control are activated.
AUTO CUT AUTO COAG 1 AUTO COAG 2 AUTO BIPOLAR
SPH Intensity no. Frequency measurement Current measurement
ATTENTION!
On the ZMK version, there is a display only during activation and if the Neurotest circuit is closed. On the
TUR version there is a display as soon as the NT LED lights up on the remote control.
SPH: The Neurotest program is activated.
Intensity no. Set intensity number is 1…5. Corresponds to the intensity number
which has been set on the remote control.
Frequency measurement Measurement of feedback frequency of the remote control.
Current measurement Measurement of feedback frequency of the output current measurement.
Setting the feedback frequency for intensity setting
(trimming potentiometer TP 1: Setting feedback frequency)
The intensity setting is converted to a frequency. This frequency can be measured at Pin 1 AD654 on IC1.
Test point MP6 is reference ground. Frequency is measured using a frequency meter. The setting takes
place at intensity 5 - for this setting the set feedback frequency is adjusted to 38.0 Hz (–0.5 Hz, +0.0 Hz) on
the frequency meter.
AUTO CUT AUTO COAG 1 AUTO COAG 2 AUTO BIPOLAR
Frequency measurement Currency [Hz]
Set Min. Max. Set Min. Max.
SPH 1 7 5 10 200 180 215
SPH 2 14 11 20 105 80 120
SPH
3
(Set value 73)
24 21 28 62 57 67
SPH 4 32 29 36 45 43 49
SPH 5 38 37 60 38 37 40