1 TEST PROGRAMS AND ADJUSTMENTS18 / 266
Test program no. 2
Call up and display of the error memory
General description
The ICC units are equipped with a system for error detection, error indication and error memory. Every error
receives an error number (ERROR no.). The unit stores the last 10 ERROR numbers. Test program 2 displays
the stored ERROR numbers. The most recent error occurring chronologically is in memory location 1.
Display
AUTO CUT AUTO COAG 1 AUTO COAG 2 AUTO BIPOLAR
Err. 1 ... 10 xxx
xxx: Display of the error number
By pressing keys 8 (Up) or 9 (Down), you can call up the memory locations one after another. By pressing
key 7, you delete the error numbers at all memory locations.
Example
After starting the test program with key 3, for example, Error no. 2 appears at memory location 1 with the
following display:
Display
AUTO CUT AUTO COAG 1 AUTO COAG 2 AUTO BIPOLAR
Err. 1 2