EasyManua.ls Logo

Erbe ICC 300 H-E - Adjustment 10 - HF Leakage Current Measurement

Erbe ICC 300 H-E
266 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 TEST PROGRAMS AND ADJUSTMENTS90 / 266
Adjustment 10
HF leakage current measurement
Procedure (This adjustment only applies to the ICC 350)
Call up Test program 16, Adjustment 10.
Connect the HF power meter (set to 350 ohms load resistance) between the active electrode of the
CUT / COAG 2 (center) socket and the potential equalization.
Jumper J3 must be plugged into the display board for this adjustment (earthed reference).
Short together the NE-ERBE original adapter cable. Activate the ICC 350 using the blue footswitch
pedal (AUTO COAG 1).
Using the power setting in the AUTO COAG 1 field, set the leakage current flowing to ground to a
value of 150 mA
eff
(tolerance 131…169 mA) (use an HF effective current meter or set a power of 7.9
watts at a HF power meter with a load resistance of 350 ohms [tolerance 6.0…10 watts (from P = I² · R
with P = 7.9 watts and R = 350 ohms is a current I = 150 mA)]).
Set TP10 on the control board (slot J3) in such a way that the HF leakage alarm is thus displayed
visually (the LED in the safety field flashes in the display indicates the value 50 in the Auto Cut
field).
Increase the leakage current by changing the power setting until an audible alarm is also emitted.
This should occur at 300…350 mA (accordingly 31.5…43 watts on a HF power meter at the above
setting).
ICC 350
AUTO CUT AUTO COAG 1 AUTO COAG 2 AUTO BIPOLAR
xxx ppp HFL Standby
• ppp Set SOFT COAG power which can be changed during activation.
• xxx Measurement value of the HF leakage current measurement.
Using the power setting in the AUTO COAG 1 field, set the HF leakage current flowing to ground at
I
eff
= 150 mA. This is not assigned for the ICC 200 and 300 since no HF leakage current monitor is
available there.

Table of Contents