EasyManua.ls Logo

Erbe ICC 300 H-E - Frequency Adjustment for Warning Tones

Erbe ICC 300 H-E
266 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1TEST PROGRAMS AND ADJUSTMENTS23 / 266
Art. No.: 80116-201
09 / 2004
Test program no. 5
Sound control for all available sounds
Frequency adjustment for warning tones
The warning tone frequency setting can be adjusted independently of other assemblies on the low-voltage
power supply (jumper J2) with Test program no. 5.
Activate Test program no. 5. MP 4 is GND for frequency counter.
Call »Tone 1«: Set frequency at MP 1 with TP 2 to 493 Hz (±2 Hz) with frequency counter.
Call »Tone 2«: Set frequency at MP 2 with TP 3 to 414 Hz (±2 Hz) with frequency counter.
Call »Tone 3«: Set frequency at MP 3 with TP 4 to 329 Hz (±2 Hz) with frequency counter.
To check settings the various tones and mixed tones are generated one after the other by calling
»Tone 4«. Here, take especial care that the operating and warning tones have different volumes.

Table of Contents