1TEST PROGRAMS AND ADJUSTMENTS9 / 266
Art. No.: 80116-201
09 / 2004
Test programs
ERBOTOM ICC 350, 300, 200
Version 4.0 / 3.0 / 2.0
No. Test program function V 4.0 / 3.0 / V 2.0
1 Basic front panel setting (only ICC 300 and 200) x
2 Calls up the Error list x
3 Test of all D-flipflop circuit memories x
4 Test of all front panel visual signals x
5 Test of all acoustic signals x
6 Test of all relays x
7 NESSY: Version number setting x
8 Display of software version no. and option no. x
9 Activation of display programs x
10 Time limit setup x
11
Measurement and display of extra-low voltages +15 volts, –15 volts,
+24 volts and the temperature
x
12 Setting the FORCED voltage (2.0: 3 x forced; 4.0: 4 x forced) x
13 not assigned
14 not assigned
15 not assigned
16 Test and setting help for all unit calibration functions x
17 Brightness setting for the seven-segment displays x
18 not assigned
19 not assigned
20 not assigned
21 not assigned
22 not assigned
23 AUTO START start delay setup x