1TEST PROGRAMS AND ADJUSTMENTS41 / 266
Art. No.: 80116-201
09 / 2004
• XX.X Voltage rating [ V ]
• ttt Temperature [°C ]
• C.aa Channel number
• yyy ADC measurement value in the range 0 ... 255
Using the “Up” (8) or “Down” (9) keys, call up the subprograms U, U-, tPt und C. 4-22.
U0 is the supply voltage controlled to +15 volts
U–1 is the supply voltage controlled to –15 volts
U2 is the supply voltage controlled to +24 volts
tPt is the temperature display of the output stage
C. 4 … 22 are the analog measurement channels according to the following table.
Voltage test U 0 or U.15
Measurement and display of the supply voltage controlled to +15 volts
Display XX.X: e.g. 14.6 equals 14.6 volts
Voltage test U–1 or –15
Measurement and display of the supply voltage controlled to –15 volts
Display XX.X: e.g. 14.6 equals –14.6 volts
Voltage test U 2 or U.24
Measurement and display of the supply voltage controlled to +24 volts
Display XX.X: e.g. 23.6 equals +23.6 volts
Measurement of the output stage temperature
Display tPt 25 means that the temperature inside the unit is 25 °C.
Test of the analog measurement channels (Subtest 4 … 22)
Measurement and display of 19 analog measurement channels inside the unit.
(Intended for internal use by the manufacturer only).
Test program no. 11
Measurement value output of the
measurement channels