Dry Ice iMiniPlus 63
3. Loss or damage caused by the ingress of moisture.
4. Evidence that the printed circuit board has been removed or tampered with.
5. Evidence of mistreatment or abuse.
6. Loss of unit.
7. Batteries are not covered by this warranty.
12. Dry Ice iMiniPlus
Important Notes:
1. Do not place the logger inside the dry ice temperatures, use external probe instead
2. Battery of iMiniPlus is not able to withstand temperatures below -40°C
3. Do not crush or cut the external probe
Specifications
Description Specifications
Product Code
MP-OE-N-8-L (Multi use External sensor)
MP-ST-N-8-L (Single use External sensor)
-100 °C to +40 °C (-148 °F to +104 °F)
Temperature Accuracy
±1.0 °C (-100 °C to -80 °C), ±1.8 °F (-148 °F to -112 °F)
±0.7 °C (-50 °C to +40 °C), ±1.3 °F (-112 °F to -58 °F)
±0.6 °C (-50 °C to +40 °C), ±1.0 °F (-58 °F to +104 °F)
External Sensor – T90 of 5 minutes in moving air
-20 °C to +70 °C (-4 °F to +158 °F)
Preprogram Option
Preprogram also available
Yes, Stop button (can be disabled)
83 x 57 x 17mm (excluding lug)
70 grams (including battery)
IP65 (with plastic cap over USB connector)
Warranty
24 months, excluding battery (Multi use loggers)
One Trip, excluding battery (Single use loggers)
ISO9001:2008, CE, RoHS, EN12830
1-2 years (depending on usage)