Eversense E3 CGM User Guide
Table 10 – Eversense E3 Calibration Stability Agreement in the PROMISE Study (continued)
Percent of CGM System Readings Within
Time
from
Calibration
Number
of Paired
CGM-YSI
Percent
15/15%
of
Reference
Percent
20/20%
of
Reference
Percent
30/30%
of
Reference
Percent
40/40%
of
Reference
Percent
> 40/40%
of
Reference
[20, 22) Hours 436 88.3 94.5 99.5 99.8 0.2
[22, 24) Hours 522 89.7 96.2 99.4 99.8 0.2
[24, 26) Hours 168 93.5 98.2 99.4 100.0 0.0
[26, 28) Hours 7 100.0 100.0 100.0 100.0 0.0
Sensor Life
Sensor life measured the percentage of sensors being able to function through the intended 180 day duration. In the
PROMISE study, 90% of sensors functioned through the 180 day period. Mean number of days was 175.
Days Sensor Life
60 98%
120 98%
180 90%
Average # of Days
175
Percentage of Sensors that lasted 60, 120 and 180 days