Finometer® Model-2 User’s Guide 23
2 Safety information
2.1 Warnings, subject safety
The data produced by Finometer® Model-2 or the accompanying
software is intended as an adjunct in subject assessment and
should not be used as a sole means for determining a subject’s
diagnosis.
Finometer® Model-2 is a
finger blood pressure monitor. Never use
the finger cuffs on other parts of the body.
Finometer® Model-2 can only be used in persons aged over 18
years.
To maintain designed operator and subject safety only use
accessories such as finger cuffs provided by Finapres Medical
Systems B.V.
Federal (USA) law restricts this device to sale by or on the order of
a physician.
Finometer® Model-2 is class B equipment according to EN 60601-
1. Use the Finometer® Model-2 only with a properly grounded AC
receptacle.
To maintain designed operator and subject safety, peripheral
equipment that is connected to Finometer® Model-2 or one of its
components, shall comply with the relevant standards.
2.2 Cautions
MRI compatibility: To prevent damage to the device, do not use
Finometer® Model-2 during Magnetic Resonance Imaging (MRI)
scanning. The cuff pressure control valve inside the frontend unit is
electromagnetically driven and cannot operate within the strong
magnetic field.
To prevent possible damage to the keys do not use sharp or hard
objects to press the keys, only use fingertips.
Never clean the device or one of its components by submersing it
into a liquid. Follow the cleaning instructions in this User’s Guide.