Finometer® Model-2 User’s Guide 47
The code ‘:’ between the beat values has a special meaning, and can
also be ‘!’ or ’?’ (see Table 4.4 Codes between the values of beat-to-
beat BP and HR).
Code Description
: Normal situation, good quality recording.
? Beat values may have become dubious, full arterial contraction is
close.
! Beat values are dubious due to arterial contraction.
Table 4.4 Codes between the values of beat-to-beat BP and HR
In Run mode when full arterial contraction is close, a typical display
looks like, e.g.:
4.2.5 Error mode
In Error mode Finometer® Model-2 has detected an error condition.
Recovery is through intervention of the operator.
A typical error message looks like:
This message indicates that an error occurred in the air supply (e.g.
pump unit, Frontend Unit air hose or finger cuff air hose). The
measurement is
stopped permanently. Press the STOP key to confirm
that the message has been noticed. After pressing the
START/STOP key,
Finometer® Model-2 switches to Ready mode.
Physiocal Height
ERROR MODE: STOP
CHECK AIR SUPPLY
Systolic Diastolic Mean Rate (bpm)
Physiocal Height
127? 84?102? 78?
CONTRCTED ARTERY
Systolic Diastolic Mean Rate (bpm)