46 Finometer® Model-2 User’s Guide
Item Description Options
127 Systolic finger pressure
To be used for trending only
84 Diastolic finger pressure To be used for trending only
102 Mean finger pressure To be used for trending only
78 Pulse rate [bpm]
-
22 Number of beats to the
next Physiocal
number in range 1..70.
OFF’ blinking when Physiocal
is switched off.
BRA Selected pressure
reconstruction
BRA, reconstructed brachial
artery pressure waveform.
FIN finger blood pressure.
C1 selected finger cuff
'c1' (C1).
+02 Current height correction
value [mmHg]
value in range –128 to +128
[mmHg].
NC’ if not connected.
NZ’ if not properly nulled.
Table 4.3 Run mode display description
CAUTION:
The precision of systolic pressure measured with the Finometer®
Model-2 does not meet the requirements of the ANSI/AAMI SP10
standard. For this reason the beat-to-beat values of Systolic,
Diastolic and Mean pressure should be used for trending only.