44 Finometer® Model-2 User’s Guide
4.2.2 Connect Frontend mode
When the Frontend Unit has not yet been connected, the display will
read, e.g.:
This display shows the message that the Frontend Unit is not
connected to the main unit. After connecting the Frontend Unit,
Finometer® Model-2 automatically switches to Ready mode.
4.2.3 Ready mode
After powering up Finometer® Model-2 with the Frontend Unit
connected, the control unit should display e.g.:
The following table describes the Ready mode display.
Item Description Options
'BRA’ Beat-to-beat blood pressure data
derived from reconstructed
brachial artery pressure
‘FIN’ Beat-to-beat blood
pressure data derived from
finger pressure
'C1’ Finger cuff connection C1
-HHH Height correction value in mmHg.
A negative sign means that the
Height Correction Unit transducer
(at the finger) is below heart
level; positive above.
value in range -128 to
+128 [mmHg].
'
NC' if not connected.
'
NZ' if not properly nulled
(zeroed).
Physiocal Height
READY MODE:START
-- BRA C1 -HHH
Systolic Diastolic Mean Rate (bpm)
Systolic Diastolic Mean Rate (bpm)
Physiocal Height
FINOMETER M2
CONNECT FRONTEND