Finometer® Model-2 User’s Guide 73
7.2.2 er cuff error messages Air pressure, fing
Error message Marker
message
Action
CHECK AIR SUPPLY The pump system fails to generate enough pre
or flow:
Check if th
ssure
e finger cuff air hose is inserted in the
Check the air hose between the Frontend Unit
and main unit.
Check for leaks or kinks.
evice checked by
Frontend Unit.
If the error persists, have the d
Finapres Medical Systems B.V.
UNACCEPTABLE P/V 64001 A pressure and/or plethysmogram error has
occurred. Pressure has been below 10 mmHg, or
gram
nt:
nger cuff is wrapped tight enough
above 250 mmHg for 2.5s, or the plethysmo
was not sufficiently small during the measureme
Retry.
Check if the fi
around the finger.
cf.
CHECK AIR SUPPLY’ and ‘CONNECT CUFF CABL’ errors.
PRESSURE LOW 64000 Mean pressure is too low. Mean pressur
below 10 mmHg for 2.5s, or too low when
perform a Physiocal.
e has been
trying to
PRESSURE TOO LOW to low during a Physiocal, the pressure
d being below 30 mmHg during
al.
64003 Pressure is
reis measu
Physioc
MEAN PRES HIGH 64002 Mean pressure is too high. Mean pressure has been
above 250 mmHg for 2.5s
UNSTABLE PRES During the pressure staircase procedure at start-up
or during a physiocal, constant pressure levels were
not stable enough:
Check if the finger cuff air hose is inserted in the
Frontend Unit.
Retry.
If the error persists, have the device checked by
Finapres Medical Systems B.V.