Finometer® Model-2 User’s Guide 53
5.2.1
T
Definition of output signals
he following outputs are available by default:
Output
Nr.
Signal Description
1 finger
Finometer® Model-2 Finger waveform is
the raw continuous finger pressure
signal.
2 reBAP
Finometer® Model-2 reBAP waveform is
the reconstructed brachial artery
pressure signal.
3 ECG input at
the front
Finometer® Model-2 ECG is the analog
ECG signal which is available when the
optional ECG module (section 3.5) is
connected to the ECG input at the front
(see Figure 1.1), or any other external
signal applied to this input.
4 Plethysmogram
Representing the volume pulsations of
the finger arteries as seen by the finger
cuff plethysmograph.
See Chapter 9 for further specifications, calibration and offset.
Note1: With BeatScope® Easy v02.10build004 it is possible to select
different signals for these 4 analog outputs. For each channel a
nals / parameters:
ial
eart Rate).
See BeatScope® Easy manual for further information.
el-2
set with
Beat-Scope® Easy, the
MARKER output will generate a voltage of
1V.
choice can be made from the following sig
Finger pressure; Height; Calibration (Square Wave), Marker*;
Finap (Filtered Finger Pressure); reBAP (reconstructed brach
artery pressure); SYS (Systolic pressure), DIA (Diastolic
pressure); MAP (Mean arterial pressure) and HR (H
*) When the MARKER key is pressed on the Finometer® Mod
keypad, the
MARKER output channel will generate a voltage of
2V as long as the key is pressed. When a marker is