Finometer® Model-2 User’s Guide 75
In this situation nothing else can be done then powering the system
nd an alarm
signal is applied by generating three beeps every minute (if the buzzer
is on, see 5.4.3), until the system is powered off.
S ta
errors.
off. The error message will stay persistent on the display a
ee the following ble for a description of possible fatal internal
Error message Description
A/D CNV NOT WELL The signal ground voltage of the Analog/Digital
converter is too high:
Power the device off and on.
If the error persists, have the device checked by
V. Finapres Medical Systems B.
PWR SUP NOT WELL A short-circuit, zero offset or power supply
oped:
ng a measurement:
.
unbalance devel
Turn off the device immediately.
Retry power-on briefly without finger cuff.
The error may also occur if you disconnect the
Frontend Unit e.g. duri
Reconnect the Frontend Unit and power the unit
off and on again.
If the error persists, have the device checked by
Finapres Medical Systems B.V
INT GND FAULT The zero offset of the pressure transducer is too
large:
Power the device off and on a number of times.
If the error persists, have the device checked by
Finapres Medical Systems B.V.