PIOX S70x 7 Connection
82 UMPIOX_S7V4-6-3EN, 2018-10-10
If transducers are replaced or added, the sensor module has to be replaced or added.
PIOX S704, S705
• Disconnect the power supply from the transmitter.
• For the connection of the AMP-Quick connector (connection system AS), see section 7.1.2.1.
• For the connection of the separate sensor module (connection system TS):
– Insert the sensor module into the lower row of terminal strip (see Fig. 7.1 or Fig. 7.2 and Fig. 7.50 or Fig. 7.51). The slots
SA1...SA4 are assigned to the transducers of measuring channel A, the slots SB1...SB4 are assigned to the transducers
of measuring channel B.
• Connect the transmitter to the power supply.
• Enter all parameters of the program branch Stop Charging.
• Press ENTER until the main menu is displayed again.
PIOX S709
• Disconnect the power supply from the transmitter.
• For the connection of the AMP-Quick connector (connection system AS), see section 7.1.2.1.
• For the connection of the separate sensor module (connection system TS):
– Insert the sensor module into the terminal strip KL5 or KL7 (see Fig. 7.3 and Fig. 7.52). The slots SA1...SA4 are as-
signed to the transducers of measuring channel A, the slots SB1...SB4 are assigned to the transducers of measuring
channel B.
• Connect the transmitter to the power supply.
• Enter all parameters of the program branch Stop Charging.
• Press ENTER until the main menu is displayed again.
Note! The serial number of sensor module and transducer have to be identical. A wrong or incorrectly con-
nected sensor module will lead to incorrect measured values or to a measurement failure.
Fig. 7.50: Transmitter PIOX S704 Fig. 7.51: Transmitter PIOX S705
Fig. 7.52: Transmitter PIOX S709
AVSAV
AGN
ARS
AR
BVSBV
BGN
BRS
BR
SA1
SA2
SA3
SA4
SB1
SB2
SB3
SB4
T1a
T1b
S2 T2a
T3a
T3b
S4
T4a
T2b
T4b
T1A
T1B
S1 T2A
T3A
T3B
S3
T4A
T2B
T4B
A+
B-
P1+ P2+
P4+
P5a
P6a P7a
P3+
101
103
P1- P2-
P4-
P5b
P6b
P7b
P3-
PE
N(-)
L(+)
sensor module
measuring channel A
sensor module
measuring channel B
AVS
AV
AGN
ARS
X2
X_BV
X_BR
X_AV
X_AR
X1
AR
BVS
BV
BGN BRS
BR
SA1 SA2 SA3 SA4 SB1
SB2
SB3
SB4
T1a T1b
S2 T2a
T3a
T3b
S4
T4a
T2b
T4b
T1A T1B
S1 T2A
T3A
T3B
S3
T4A
T2B
T4B
A+
B-
P1+ P2+
P4+
P5a
P6a P7a
P3+
101
103
P1- P2-
P4-
P5b
P6b
P7b
P3-
PE
N(-)
L(+)
sensor module
measuring channel A
sensor module
measuring channel B
K L 7
K L 5
K L 3
K L 1
X 7
K L 4K L 2
X 6 A V
X 8 B V
X 6 A R X 8 B R
X 5
BA
K L 6
K L 8
C H A N N E L
S A 3
S A 1
S A 2
S A 4
S B 3
S B 1
S B 2
S B 4
A V S
A V
A R S
A G N
A R
B V S
B V
B R S
B G N
B R
4 A +
4 B -
4 2
4 3
4 1
L +
L -
N
L 1
P E
P 6 a
P 7 a
P 7 +
P 5 a
P 6 +
P 2 +
P 5 +
P 1 +
P 3 +
P 4 +
P 6 b
P 7 b
P 7 -
P 5 b
P 6 -
P 2 -
P 5 -
P 1 -
P 3 -
P 4 -
T 4 A
T 4 B
T 3 B
S 3
T 3 A
T 1 B
T 2 B
T 1 A
S 1
T 2 A
T 4 a
T 4 b
T 3 b
S 4
T 3 a
T 1 b
T 2 b
T 1 a
S 2
T 2 a
sensor module
measuring channel A
sensor module
measuring channel B

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim PIOX S704 NN Series and is the answer not in the manual?

Flexim PIOX S704 NN Series Specifications

General IconGeneral
BrandFlexim
ModelPIOX S704 NN Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals