Thermographic measurement techniques35
5. Measure the apparent temperature of the aluminum foil and write it down. The foil is
considered a perfect reflector, so its apparent temperature equals the reflected appa-
rent temperature from the surroundings.
Figure 35.5 Measuring the apparent temperature of the aluminum foil.
35.2.1.2 Step 2: Determining the emissivity
Follow this procedure:
1. Select a place to put the sample.
2. Determine and set reflected apparent temperature according to the previous
procedure.
3. Put a piece of electrical tape with known high emissivity on the sample.
4. Heat the sample at least 20 K above room temperature. Heating must be reasonably
even.
5. Focus and auto-adjust the camera, and freeze the image.
6. Adjust Level and Span for best image brightness and contrast.
7. Set emissivity to that of the tape (usually 0.97).
8. Measure the temperature of the tape using one of the following measurement
functions:
Isotherm (helps you to determine both the temperature and how evenly you have
heated the sample)
Spot (simpler)
Box Avg (good for surfaces with varying emissivity).
9. Write down the temperature.
10. Move your measurement function to the sample surface.
11. Change the emissivity setting until you read the same temperature as your previous
measurement.
12. Write down the emissivity.
Note
Avoid forced convection
Look for a thermally stable surrounding that will not generate spot reflections
Use high quality tape that you know is not transparent, and has a high emissivity you
are certain of
This method assumes that the temperature of your tape and the sample surface are
the same. If they are not, your emissivity measurement will be wrong.
#T559954; r. AP/42311/42335; en-US
214

Table of Contents

Other manuals for FLIR T1020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the FLIR T1020 and is the answer not in the manual?

FLIR T1020 Specifications

General IconGeneral
Resolution1024 × 768 pixels
Detector TypeUncooled microbolometer
Digital Camera5 MP
Laser PointerYes
Wi-FiYes
Image Frequency30 Hz
Detector Pitch17 µm
Digital Zoom1-8x continuous
Video RecordingYes
Storage MediaSD card
Spectral Range7.5 – 14 µm
Temperature Range-40 °C to +2000 °C (-40 °F to +3632 °F)
Field of View28° x 21°
Display4.3 in LCD touchscreen, 800 × 480 pixels
FocusMotorized and manual
Image ModesThermal, visual, MSX
Operating Temperature-15 °C to +50 °C
Storage Temperature-40 °C to +70 °C (-40 °F to +158 °F)
ConnectivityUSB 2.0, HDMI
Image StorageJPEG
Image FormatsJPEG
Thermal Sensitivity<20 mK at 30°C

Related product manuals