Power Recorder
Setting Up the Recorder
47
default directory My Documents\Fluke\PQ, a dialog box opens so that you can
navigate to the file.
Note
If clicking on an annotation marker does not seem to do anything, the Zoom
box is probably selected in the graph toolbar, instead of the Select tool.
Click the Select tool and then select the annotation.
To see overlapping annotations:
Sometimes annotations are so close to each other that they actually overlap on the
Volts/Amps/Hz reference graph. When this is the case, if you click on overlapping
annotations, a menu opens and you can select either Open, to open the annotation now on
top, or Select Next Overlapping Annotation, to bring the next one to the top.
To open the annotation now on top, choose Open.
To bring the next annotation to the top, choose Select Next Overlapping Annotation.
To then open that annotation, click on it again and choose Open.
Note
To see overlapping annotations more clearly on the Volts/Amps/Hz
reference graph, move the selection cursors so that they just surround the
area that contains the overlapping illustrations. The Detail graph adjusts
as the cursors move, zooming in to show the area between the selection
cursors. You can now see the annotations separately.
Using Image or Voice Annotations
While you are monitoring data with a Recorder, you may find it helpful to take a picture
that relates to the power quality data you are recording, or to record verbal comments as
you observe the conditions or review the data being measured. You can then associate the
image or recording with the data being recorded by adding annotations that reference
them.
The PDA includes a microphone for recording notes which are saved as .WAV files, and
an optional camera accessory is available on the PDA which will capture images as .jpg
files.
You can associate both images and audio files with the data being gathered in the
Recorder by inserting annotations and specifying the name of a .WAV or .jpg file. After
you download data from the Recorder, you can then move the image and audio files to
the PC along with the recorded data (.odn) file.
Note
You can also associate other types of files with the annotation. For example,
if you had a .MPG file with a movie, or a .DOC file with associated
information. You can enter any file name using this annotation method.
Inserting Image or Voice Annotations
1. Set up the Recorder and begin monitoring.
2. Using your PDA-cam (not available in standard PDA supplied) or another camera,
take a picture of what you want to capture.
3. Click the Annotate icon to display the Annotate screen.
4. Move the image or audio file to a location that will make it easier to Transfer it to a
PC later, when you download and move the recorded data file to your PC. Place the

Table of Contents

Other manuals for Fluke 1750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1750 and is the answer not in the manual?

Fluke 1750 Specifications

General IconGeneral
TypePower Quality Analyzer
Frequency Measurement Range42.5 Hz to 69 Hz
Safety RatingCAT IV 600 V, CAT III 1000 V
Accuracy±0.1% of reading
HarmonicsUp to 50th harmonic
CommunicationEthernet, USB
Power Supply100 V to 240 V AC
Channels4 voltage, 4 current
Battery BackupYes, internal battery
Measurement ParametersVoltage, Current, Power, Harmonics, Flicker
Data Logging Interval1 second to 60 minutes
Battery Life45 minutes (UPS)
Voltage Measurement Range1 to 1000 V (phase to neutral)
Operating Temperature-10 to 50 °C